An experimental approach to efficiency calibration for gamma-ray spectrometric analysis of large air particulate filters

2013

Abstract A full-energy-peak efficiency (FEPE) calibration procedure for gamma-ray spectrometric analysis of air particulate samples collected on large filters is described herein. The experimental results are obtained for an unconventional measurement geometry, termed a “packet-sample”. The sample is obtained from a large cellulose filter (45 cm×45 cm) used to collect air particulate samples that is resized to dimensions suitable for spectrometric measurements (6 cm×6 cm×0.7 cm). To determine the FEPEs, many standards were created, i.e., some filters containing a small amount of ThO 2 and others containing a known amount of KCl. Efficiency curves obtained through best fits to experimental d…

RadiationChemistrySettore ING-IND/20 - Misure E Strumentazione NucleariDetectorGamma rayAnalytical chemistryMineralogyExperimental dataEfficiency calibration Air particulate Gamma-ray spectrometry HPGe detectorsParticulatesSemiconductor detectorFilter (large eddy simulation)CalibrationGamma ray spectrometry